MITEx: A Portable Hand Exoskeleton for Assessment and Treatment in Neurological Rehabilitation

IEEE Int Conf Rehabil Robot. 2023 Sep:2023:1-6. doi: 10.1109/ICORR58425.2023.10304721.

Abstract

This work describes the design and preliminary characterization of a novel portable hand exoskeleton for poststroke rehabilitation. The platform actively mobilizes the index-metacarpophalangeal (I-MCP) joint, and it additionally offers individual rigid support to distal degrees of freedom (DoFs) of the index and thumb. The test-bench characterization proves the capability of the device to render torques at the I-MCP level with high fidelity within frequencies of interest for the application (up to 3 Hz). The introduction of a feed-forward friction compensation at the actuator level lowers the output mechanical stiffness by 32%, contributing to a highly transparent behavior; moreover, the functionality of the platform in rendering different interaction strategies (patient/robot-in-charge) is tested with three healthy subjects, showing the potential of the device to provide assistance as needed.

MeSH terms

  • Exoskeleton Device*
  • Hand
  • Humans
  • Metacarpophalangeal Joint
  • Neurological Rehabilitation*
  • Thumb