Abstract

Aims: Transvenous lead extraction (TLE) is important in the management of cardiac implantable electronic devices but carries risk. It is most commonly completed from the superior access, often with 'bail-out' support via the femoral approach. Superior and inferior access may be used in tandem, which has been proposed as an advance in safety and efficacy. The aim of this study is to evaluate the safety and efficacy of the Tandem approach.

Method: The 'Tandem' procedure entailed grasping of the targeted lead in the right atrium to provide countertraction as a rotational dissecting sheath was advanced over the lead from the subclavian access. Consecutive 'Tandem' procedures performed by a single operator between December 2020 and March 2023 in a single large-volume TLE centre were included and compared with the conventional superior approach (control) using 1:1 propensity score matching; patients were statistically matched for demographics.

Results: The Tandem in comparison with the conventional approach extracted leads of much greater dwell time (148.9 ± 79 vs. 108.6 ± 77 months, P < 0.01) in a shorter procedure duration (96 ± 36 vs. 127 ± 67 min, P < 0.01) but requiring more fluoroscopy (16.4 ± 10.9 vs. 10.8 ± 14.9 min, P < 0.01). The Tandem and control groups had similar clinical (100% vs. 94.7%, P = 0.07) and complete (94.8% vs. 92.8%, P = 0.42) success, with comparable minor (4% vs. 6.7%, P = 0.72) and major (0% vs. 4%, P = 0.25) complications; procedural (0% vs. 1.3%, P = 1) and 30-day (1.3% vs. 4%, P = 0.62) mortality were also similar.

Conclusion: The Tandem procedure is as safe and effective as the conventional TLE. It can be applied to leads of a long dwell time with a potentially shorter procedure duration.

Keywords: Femoral extraction; Lead extraction; Non-laser transvenous lead extraction; Pacemaker extraction; Tandem approach; Transvenous lead extraction.

MeSH terms

  • Defibrillators, Implantable* / adverse effects
  • Device Removal / adverse effects
  • Device Removal / methods
  • Fluoroscopy
  • Humans
  • Pacemaker, Artificial* / adverse effects
  • Retrospective Studies
  • Time Factors
  • Treatment Outcome

Grants and funding