Factors Affecting Diagnostic Yield of Lesional Bone Biopsy in Hematologic Malignancy Patients

Cardiovasc Intervent Radiol. 2024 Jan;47(1):80-86. doi: 10.1007/s00270-023-03594-9. Epub 2023 Nov 1.

Abstract

Purpose: To evaluate factors affecting the diagnostic yield (percent of biopsy samples leading to a pathologic diagnosis) of lesional bone biopsies in patients with hematologic malignancies.

Materials and methods: This retrospective study included 206 lesional bone biopsies in 182 patients with a hematologic malignancy between January 2017 and December 2022. The parameters that were reviewed to evaluate diagnostic yield included biopsy device type (manual vs. electric-powered drill), number of biopsy cores acquired, core biopsy needle gauge, preliminary intra-procedural sample adequacy (touch preparation cytology determining if samples are adequate for final pathologic examination), lesion morphology on Computed Tomography (CT), and presence of crush artifact.

Results: Review of 206 lesional biopsies showed overall diagnostic yield to be 89.8% (185/206). The two statistically significant factors affecting diagnostic yield were biopsy device type and in-room adequacy. 41/42 samples obtained with the electric-powered drill and 144/164 samples obtained using a variety of manual needles were diagnostic (97.6% vs 87.8%, p = 0.03). Of the 192 samples that were assessed for sample adequacy intra-procedurally, 97/102 of the samples that were deemed adequate were diagnostic, and 77/90 of the samples where intra-procedural adequacy was not confirmed were diagnostic (95.1% vs 85.6%, p = 0.018). The remaining factors did not affect diagnostic yield.

Conclusion: The use of an electric-powered drill bone biopsy device and intra-procedural confirmation of sample adequacy are associated with a higher diagnostic yield of lesional bone biopsies in patients with hematologic malignancies. The presence or absence of crush artifact did not significantly affect the diagnostic yield in these patients.

Keywords: Bone biopsy; CT-guided; Diagnostic yield; Drill system; Hematologic malignancy.

MeSH terms

  • Biopsy, Fine-Needle
  • Bone Diseases*
  • Bone and Bones / pathology
  • Hematologic Neoplasms*
  • Humans
  • Image-Guided Biopsy / methods
  • Retrospective Studies