Abstract

Background: Drug-coated balloons (DCB) are an emerging tool for modern percutaneous coronary intervention (PCI), but evidence on their use for de novo lesions on large vessels is limited.

Methods: Consecutive patients undergoing DCB-based PCI on the left anterior descending artery in 2 Italian centers from 2018 to 2022 were retrospectively enrolled and compared with patients who received left anterior descending PCI with contemporary drug-eluting stents (DES). In-stent restenosis was excluded. The DCB group included both patients undergoing DCB-only PCI and those receiving hybrid PCI with DCB and DES combined. The primary end point was target lesion failure at 2 years, defined as the composite of target lesion revascularization, cardiac death, and target vessel myocardial infarction.

Results: We included 147 consecutive patients undergoing DCB-based treatment on the left anterior descending artery and compared them to 701 patients who received conventional PCI with DES. In the DCB group, 43 patients (29.2%) were treated with DCB only and 104 (70.8%) with a hybrid approach; DCB length was greater than stent length in 55.1% of cases. Total treated length was higher in the DCB group (65 [40-82] versus 56 [46-66] mm; P=0.002), while longer DESs were implanted (38 [24-62] versus 56 [46-66] mm; P<0.001) and a higher rate of large vessels were treated (76.2% versus 83.5%; P=0.036) in the DES cohort. The cumulative 2-year target lesion failure incidence was not significantly different between the 2 groups (DCB, 4.1% versus DES, 9.8%; hazard ratio, 0.51 [95% CI, 0.20-1.27]; P=0.15). After a 1:1 propensity score matching resulting in 139 matched pairs, the DCB-based treatment was associated with a lower risk for target lesion failure at 2 years compared with DES-only PCI (hazard ratio, 0.2 [95% CI, 0.07-0.58]; P=0.003), mainly driven by less target lesion revascularization.

Conclusions: A DCB-based treatment approach for left anterior descending revascularization allows a significantly reduced stent burden, thereby potentially limiting target lesion failure risk at midterm follow-up.

Keywords: balloon angioplasty; coronary restenosis; drug-eluting stents; propensity score.

MeSH terms

  • Angioplasty, Balloon, Coronary* / adverse effects
  • Angioplasty, Balloon, Coronary* / methods
  • Coated Materials, Biocompatible
  • Coronary Artery Disease* / complications
  • Coronary Artery Disease* / diagnostic imaging
  • Coronary Artery Disease* / therapy
  • Coronary Restenosis* / diagnostic imaging
  • Coronary Restenosis* / etiology
  • Coronary Restenosis* / therapy
  • Coronary Vessels / diagnostic imaging
  • Drug-Eluting Stents* / adverse effects
  • Humans
  • Percutaneous Coronary Intervention* / adverse effects
  • Retrospective Studies
  • Treatment Outcome

Substances

  • Coated Materials, Biocompatible