Correct positioning of the calcar screw leads to superior results in proximal humerus fractures treated with carbon-fibre-reinforced polyetheretherketone plate osteosynthesis with polyaxial locking screws

J Orthop Traumatol. 2023 Oct 10;24(1):54. doi: 10.1186/s10195-023-00733-3.

Abstract

Background: Plate osteosynthesis with implants made of carbon-fibre-reinforced polyetheretherketone (CFR-PEEK) has recently been introduced for the treatment of fractures of the proximal humerus (PHFs). The advantages of the CFR-PEEK plate are considered to be its radiolucency, its favourable modulus of elasticity, and the polyaxial placement of the screws with high variability of the angle. The primary aim of this study is to investigate the influence of calcar screw positioning on the complication and revision rates after CFR-PEEK plating of PHFs. The secondary aim is to assess its influence on functional outcome.

Material and methods: Patients were identified retrospectively. Minimum follow-up was 12 months. The cohort was divided into two groups depending on the distance of the calcar screw to the calcar (group I: < 12 mm, group II: ≥ 12 mm). The range of motion (ROM), Subjective Shoulder Value Score (SSV) and Constant-Murley Score (CS) were analysed at follow-up examination. Subjective complaints, complications (e.g. humeral head necrosis, varus dislocation) and the revision rate were evaluated.

Results: 51 patients (33 female, 18 male) with an average age of 68.6 years were included after a period of 26.6 months (group I: 32 patients, group II: 19 patients). Apart from the gender distribution, no significant differences were seen in the patient characteristics. The outcome scores showed significantly better clinical results in group I: SSV 83.4 vs 71.2, p = 0.007; CS 79.1 vs 67.8, p = 0.013. Complications were seen less frequently in group I (18.8 % vs 47.4 %, p = 0.030).

Conclusion: This study shows that the positioning of the calcar screw is relevant for CFR-PEEK plate osteosynthesis in PHFs with a good reduction of the fracture. Optimal positioning of the calcar screw close to the calcar (< 12 mm) is associated with a lower rate of complications, resulting in significantly superior functional outcomes.

Level of evidence: III, retrospective cohort study.

Keywords: Angular stable plate osteosynthesis; Avascular necrosis; Constant–Murley score; PEEK; Subjective shoulder value; Varus dislocation.

MeSH terms

  • Aged
  • Bone Plates / adverse effects
  • Bone Screws / adverse effects
  • Carbon Fiber
  • Female
  • Fracture Fixation, Internal / methods
  • Humans
  • Humeral Fractures*
  • Humerus / surgery
  • Ketones
  • Male
  • Polyethylene Glycols
  • Retrospective Studies
  • Shoulder Fractures* / diagnosis
  • Shoulder Fractures* / surgery
  • Treatment Outcome

Substances

  • polyetheretherketone
  • Carbon Fiber
  • Polyethylene Glycols
  • Ketones