Virtual implantation of Derivo Embolization Devices using PreSize Neurovascular software: Accuracy evaluation and size selection comparison study

Interv Neuroradiol. 2023 Oct 8:15910199231202272. doi: 10.1177/15910199231202272. Online ahead of print.

Abstract

Background: Evaluating the deployed length of flow diverting stents (FDs) to select the optimal device size remains a challenging, yet crucial, task in aneurysm treatment. This study reports on the accuracy of PreSize Neurovascular (Oxford Heartbeat Ltd), a visualization and simulation software for FD intervention planning, in predicting FD deployed length, and on its impact on device size selection.

Methods: Imaging data from consecutive patients treated with Derivo Embolization Device (Acandis GmbH) were collected from University Medical Center Hamburg-Eppendorf and retrospectively analyzed. Accuracy evaluation: prediction accuracy was calculated by comparing deployed FD lengths measured from imaging data and simulated by PreSize. Size selection comparison: two Interventional Neuroradiologists (INR1, INR2), blinded to the devices deployed, used PreSize to select the optimal device size (diameter and length). Their choices were compared against the deployed devices selected by conventional planning.

Results: Among 98 implanted devices, PreSize predicted deployed FD length with a mean accuracy of 94.54% (95% confidence interval [93.72%, 95.35%]). Among 98 aneurysm cases, PreSize-informed device lengths were significantly shorter (Wilcoxon signed-rank test, INR1: W = 394, P < .001, INR2: W = 305, P < .001) by 4.13 and 4.18 mm on average, and up to 20 and 25 mm, for INR1 and INR2, respectively, than the conventionally selected FDs. In 32% of cases, PreSize-informed devices resulted in fewer vessel bends covered by the FD while achieving sufficient aneurysm coverage.

Conclusions: PreSize retrospectively predicted deployed FD lengths with high accuracy. Moreover, INRs in this study were more inclined to select shorter stent length in the simulation than they would have done conventionally.

Keywords: Device selection; aneurysm treatment; flow diverter sizing; intervention planning; simulation software.