[Research of implant accuracy using a digital registration method]

Shanghai Kou Qiang Yi Xue. 2023 Jun;32(3):287-291.
[Article in Chinese]

Abstract

Purpose: The aim of this study was to introduce a new method to evaluate the clinical accuracy of implant position. The results were compared to traditional cone beam CT (CBCT) method.

Methods: A total of 36 implants from 24 patients with sufficient bone volume were enrolled into the study. CBCT method and digital registration method were compared to evaluate the accuracy of implant position. The measurement parameters were defined as deviations between ideal and postsurgical implant position at occlusal point(d1), apical point(d2) and axis(α). The deviations between two methods were analyzed with SPSS 19.0 software package.

Results: The deviations between ideal and postsurgical implant position using CBCT were (0.88±0.64) mm for occlusal point, (1.07±0.85) mm for apical point and (4.74±2.35)° for angle. In digital registration method, the deviations were (0.86±0.67) mm for occlusal point, (1.12±0.88) mm for apical point and (4.56±2.66)° for angle. No significant difference(P>0.05) was found between the two methods.

Conclusions: There was no significant difference between the two methods in evaluating the clinical accuracy of implant position. Digital registration method could be accepted in clinical application.

Publication types

  • English Abstract

MeSH terms

  • Computer-Aided Design
  • Cone-Beam Computed Tomography / methods
  • Dental Implantation, Endosseous / methods
  • Dental Implants*
  • Humans
  • Imaging, Three-Dimensional
  • Surgery, Computer-Assisted* / methods

Substances

  • Dental Implants