Evaluation of scanning accuracy for two commercially available intraoral scanners in reproducing orthodontic bracket dimensions

Eur Rev Med Pharmacol Sci. 2023 Sep;27(17):7898-7906. doi: 10.26355/eurrev_202309_33548.

Abstract

Objective: The aim of this study was to compare the scanning accuracy of two different intra-oral scanners- MEDIT i500 and TRIOS 3 shape in reproducing orthodontic bracket dimensions.

Materials and methods: This in vivo cross-sectional study comprised seven subjects with a full complement of permanent dentition without third molars. Complete arch scanning was carried out with two intraoral scanners, such as MEDIT i500, TRIOS 3 shape, after bonding with brackets. The control group consisted of bracket dimensions measured directly by using Vernier calipers before bonding. Bracket dimensions of three-dimensional (3D) images were measured by using OrthoAnalyzer software. The accuracy of intraoral scanning was investigated by comparing bracket dimensions among the three groups using One-way ANOVA and Post-Hoc Tukey HSD test, and by evaluating outcomes for each quadrant and an individual tooth in complete-arch scans.

Results: When comparing bracket dimensions of 3D images with manual measurements using a traditional Vernier caliper, MEDIT i500 showed no significant difference when compared to the control group (p>0.05) in full arch scanning as well as the quadrant and single tooth scans in complete arch scanning. TRIOS 3 shape showed a significant difference when compared to the control and MEDIT i500 group (p<0.05) for all parameters.

Conclusions: MEDIT i500 showed higher accuracy and better reproduction of orthodontic bracket dimensions than TRIOS 3 shape.

MeSH terms

  • Cross-Sectional Studies
  • Humans
  • Molar, Third
  • Object Attachment
  • Orthodontic Brackets*
  • Radionuclide Imaging