Purpose: Studies of flash glucose monitoring systems (FGMSs) in pregnancy are insufficient, especially in gestational diabetes (GD). Our aim was to evaluate Freestyle Libre's usability and accuracy (compared to self-monitoring of blood glucose [SMBG]) for GD patients in real-life conditions.
Methods: This is a prospective study with pregnant women diagnosed with GD (n = 24 for the usability analysis; n = 19 for the accuracy analysis). The study duration was up to 28 days (lifetime of two sensors). Participants executed a minimum of four daily FGMS readings obtained immediately after capillary SMBG. Analytical accuracy was assessed with mean absolute relative difference (MARD) and mean absolute difference (MAD); clinical accuracy was assessed with Surveillance Error Grid (SEG). Usability was evaluated with a user acceptability questionnaire.
Results: The mean pregestational BMI was 25.21 ± 5.15 kg/m2 (mean ± SD), the mean gestational age was 30.31 ± 2.02 weeks, and the mean glucose values were 76.63 ± 7.49 mg/dL. A total of 1339 SMBG-FGMS pairs of values were obtained. Analytical accuracy was good with an overall MARD of 14.07% and an in-target MARD of 13.79%. The number of SMBG-FMGS pairs for above-target values was low (122 of 1339) with an associated MARD of 17.95%. Clinical accuracy of the FGMS was demonstrated, with 94.4% of values in the no-risk or slight, lower risk zones of the SEG. FGMS accuracy was unaffected by pregestational BMI or gestational age. The user acceptability questionnaire showed high levels of satisfaction, with 95.8-100% preferring FGMS to SMBG. No unexpected or severe adverse effects occurred.
Conclusion: FGMS showed good performance in GD regarding accuracy and usability. Larger studies are needed to corroborate our results, verify the analytical accuracy of above-target values as this glucose range might lead to initiation or adjustment of pharmacological therapy, and ultimately establish definitive recommendations regarding prescription of FGMS for GD patients.
Keywords: Accuracy; Flash glucose monitoring system; Gestational diabetes; Usability.
© 2023. The Author(s), under exclusive licence to Hellenic Endocrine Society.