Real-time hybrid angular-interrogation surface plasmon resonance sensor in the near-infrared region for wide dynamic range refractive index sensing

Sci Rep. 2023 Sep 20;13(1):15655. doi: 10.1038/s41598-023-42873-w.

Abstract

Herein, we integrated angle-scanning surface plasmon resonance (SPR) and angle-fixed SPR as a hybrid angular-interrogation SPR to enhance the sensing performance. Galvanometer-mirror-based beam angle scanning achieves a 100-Hz acquisition rate of both the angular SPR reflectance spectrum and the angle-fixed SPR reflectance, whereas the use of near-infrared light enhances the refractive index (RI) sensitivity, range, and precision compared with visible light. Simultaneous measurement of the angular SPR reflectance spectrum and angle-fixed SPR reflectance boosts the RI change range, RI resolution, and RI accuracy to 10-1-10-6 RIU, 2.24 × 10-6 RIU, and 5.22 × 10-6 RIU, respectively. The proposed hybrid SPR is a powerful tool for wide-dynamic-range RI sensing with various applications.