Performance of a multi-sensor implantable defibrillator algorithm for heart failure monitoring in the presence of atrial fibrillation

Europace. 2023 Aug 2;25(9):euad261. doi: 10.1093/europace/euad261.

Abstract

Aims: The HeartLogic Index combines data from multiple implantable cardioverter defibrillators (ICDs) sensors and has been shown to accurately stratify patients at risk of heart failure (HF) events. We evaluated and compared the performance of this algorithm during sinus rhythm and during long-lasting atrial fibrillation (AF).

Methods and results: HeartLogic was activated in 568 ICD patients from 26 centres. We found periods of ≥30 consecutive days with an atrial high-rate episode (AHRE) burden <1 h/day and periods with an AHRE burden ≥20 h/day. We then identified patients who met both criteria during the follow-up (AHRE group, n = 53), to allow pairwise comparison of periods. For control purposes, we identified patients with an AHRE burden <1 h throughout their follow-up and implemented 2:1 propensity score matching vs. the AHRE group (matched non-AHRE group, n = 106). In the AHRE group, the rate of alerts was 1.2 [95% confidence interval (CI): 1.0-1.5]/patient-year during periods with an AHRE burden <1 h/day and 2.0 (95% CI: 1.5-2.6)/patient-year during periods with an AHRE-burden ≥20 h/day (P = 0.004). The rate of HF hospitalizations was 0.34 (95% CI: 0.15-0.69)/patient-year during IN-alert periods and 0.06 (95% CI: 0.02-0.14)/patient-year during OUT-of-alert periods (P < 0.001). The IN/OUT-of-alert state incidence rate ratio of HF hospitalizations was 8.59 (95% CI: 1.67-55.31) during periods with an AHRE burden <1 h/day and 2.70 (95% CI: 1.01-28.33) during periods with an AHRE burden ≥20 h/day. In the matched non-AHRE group, the rate of HF hospitalizations was 0.29 (95% CI: 0.12-0.60)/patient-year during IN-alert periods and 0.04 (95% CI: 0.02-0.08)/patient-year during OUT-of-alert periods (P < 0.001). The incidence rate ratio was 7.11 (95% CI: 2.19-22.44).

Conclusion: Patients received more alerts during periods of AF. The ability of the algorithm to identify increased risk of HF events was confirmed during AF, despite a lower IN/OUT-of-alert incidence rate ratio in comparison with non-AF periods and non-AF patients.

Clinical trial registration: http://clinicaltrials.gov/Identifier: NCT02275637.

Keywords: Atrial fibrillation; CRT; Heart failure; ICD; Risk stratification.

Publication types

  • Clinical Study

MeSH terms

  • Algorithms
  • Atrial Fibrillation* / diagnosis
  • Atrial Fibrillation* / epidemiology
  • Atrial Fibrillation* / therapy
  • Defibrillators, Implantable*
  • Heart Atria
  • Heart Failure* / diagnosis
  • Heart Failure* / epidemiology
  • Heart Failure* / therapy
  • Humans

Associated data

  • ClinicalTrials.gov/NCT02275637