Comparative Study of Ablation Zone of EMPRINT HP Microwave Device with Contemporary 2.4 GHz Microwave Devices in an Ex Vivo Porcine Liver Model

Diagnostics (Basel). 2023 Aug 18;13(16):2702. doi: 10.3390/diagnostics13162702.

Abstract

(1) Background: Percutaneous microwave ablation (MWA) is an accepted treatment of non-operative liver cancer. This study compares the ablation zones of four commercially available 2.45 GHz MWA systems (Emprint, Eco, Neuwave, and Solero) in an ex vivo porcine liver model. (2) Methods: Ex vivo porcine livers (n = 85) were obtained. Two ablation time setting protocols were evaluated, the manufacturer's recommended maximum time and a 3 min time, performed at the manufacturer-recommended maximum power setting. A total of 236 ablation samples were created with 32 (13.6%) samples rejected. A total of 204 samples were included in the statistical analysis. (3) Results: For single-probe protocols, Emprint achieved ablation zones with the largest SAD. Significant differences were found in all comparisons for the 3 min time setting and for all comparisons at the 10 min time setting except versus Neuwave LK15 and Eco. Emprint produced ablation zones that were also significantly more spherical (highest SI) than the single-probe ablations from all other manufacturers. No statistical differences were found for ablation shape or SAD between the single-probe protocols for Emprint and the three-probe protocols for Neuwave. (4) Conclusions: The new Emprint HP system achieved large and spherical ablation zones relative to other 2.45 GHz MWA systems.

Keywords: ex vivo bovine liver; hepatocellular carcinoma; liver tumours; microwave ablation; percutaneous thermal ablation.

Grants and funding