Comparison of Novel Volumetric Microperimetry Metrics in Intermediate Age-Related Macular Degeneration: PINNACLE Study Report 3

Transl Vis Sci Technol. 2023 Aug 1;12(8):21. doi: 10.1167/tvst.12.8.21.

Abstract

Purpose: To investigate and compare novel volumetric microperimetry (MP)-derived metrics in intermediate age-related macular degeneration (iAMD), as current MP metrics show high variability and low sensitivity.

Methods: This is a cross-sectional analysis of microperimetry baseline data from the multicenter, prospective PINNACLE study (ClinicalTrials.gov NCT04269304). The Visual Field Modeling and Analysis (VFMA) software and an open-source implementation (OSI) were applied to calculate MP-derived hill-of-vison (HOV) surface plots and the total volume (VTOT) beneath the plots. Bland-Altman plots were used for methodologic comparison, and the association of retinal sensitivity metrics with explanatory variables was tested with mixed-effects models.

Results: In total, 247 eyes of 189 participants (75 ± 7.3 years) were included in the analysis. The VTOT output of VFMA and OSI exhibited a significant difference (P < 0.0001). VFMA yielded slightly higher coefficients of determination than OSI and mean sensitivity (MS) in univariable and multivariable modeling, for example, in association with low-luminance visual acuity (LLVA) (marginal R2/conditional R2: VFMA 0.171/0.771, OSI 0.162/0.765, MS 0.133/0.755). In the multivariable analysis, LLVA was the only demonstrable predictor of VFMA VTOT (t-value, P-value: -7.5, <0.001) and MS (-6.5, <0.001).

Conclusions: The HOV-derived metric of VTOT exhibits favorable characteristics compared to MS in evaluating retinal sensitivity. The output of VFMA and OSI is not exactly interchangeable in this cross-sectional analysis. Longitudinal analysis is necessary to assess their performance in ability-to-detect change.

Translational relevance: This study explores new volumetric MP endpoints for future application in therapeutic trials in iAMD and reports specific characteristics of the available HOV software applications.