Dependence of ghost on the incident light angle into dichroic mirror

J Biophotonics. 2023 Nov;16(11):e202300190. doi: 10.1002/jbio.202300190. Epub 2023 Aug 18.

Abstract

The dichroic mirror (DM) is a key component in microscope. We found a ghost in the reflection channel of a dual-channel fluorescence microscope and studied the relationship between the ghost and the incidence angle θ into the DM. The DM emission surface reflection generated ghost if the θ is not 45 ° . We analyzed the distance and intensity relationship between the ghost and the primary image, which is θ -dependent and was demonstrated by imaging live cells and a stage micrometer. The ghost can be eliminated by placing the DM between objective and tube lens, but not between tube lens and detector, ensuring that the incident light into the DM is approximately parallel. Furthermore, the transmitted light of the DM is shifted towards a longer wavelength with increasing θ . Collectively, microscopists must carefully optimize the θ when designing a microscope to avoid the ghost.

Keywords: dichroic mirror; incidence angle; microscopic imaging; optical microscope.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Microscopy, Fluorescence* / methods