Outcomes of toric intraocular lens realignment surgery done using slit-lamp method and wavefront aberrometry method

Indian J Ophthalmol. 2023 Aug;71(8):2978-2983. doi: 10.4103/IJO.IJO_37_23.

Abstract

Purpose: To compare the slit-lamp method and wavefront aberrometry method based on outcomes of toric realignment surgeries.

Settings: Tertiary care ophthalmic hospital.

Design: Retrospective study.

Methods: This study included all eyes undergoing toric intraocular lens (TIOL) realignment surgery between January 2019 and December 2021 for which TIOL axis assessment by slit-lamp method and wavefront aberrometry method was available. Data were retrieved from electronic medical records, and we documented demographics, uncorrected visual acuity (UCVA), subjective refraction, and TIOL axis by slit-lamp and wavefront aberrometry methods on postoperative day 1 and day 14. In patients with misalignment, TIOL was realigned to the original position in group 1 (27 patients) and to an axis based on calculations provided by wavefront aberrometer in group 2 (25 patients). Post-realignment surgery, UCVA, subjective refraction, and TIOL axis by slit-lamp and wavefront aberrometry methods were assessed and analyzed.

Results: We analyzed 52 eyes and found that the mean preoperative misalignment with the slit-lamp method (44.9° ±20.0°) and wavefront aberrometry (47.1° ±19.5°) was similar. The corresponding degrees of misalignment post-TIOL repositioning surgeries were 5.2° ±5.2° (slit-lamp method) and 4.7° ±5.1° (wavefront aberrometry) (P = 0.615). Both groups showed significant improvement in median log of minimum angle of resolution (logMAR) UCVA and reduction in median refractive cylinder.

Conclusions: Slit-lamp method is as good as wavefront aberrometer method to assess TIOL axis. Toric realignment surgery is found to be safe, and realigning TIOL based on either slit-lamp method or wavefront aberrometer method equally improved UCVA and decreased residual refractive cylinder.

Keywords: Misaligned toric IOL; realignment surgery; slit-lamp method; toric IOL; wavefront aberrometer.