Development of an automatic device performing chest compression and external defibrillation: An animal-based pilot study

PLoS One. 2023 Jul 26;18(7):e0288688. doi: 10.1371/journal.pone.0288688. eCollection 2023.

Abstract

Background: Automatic chest compression devices (ACCDs) can promote high-quality cardiopulmonary resuscitation (CPR) and are widely used worldwide. Early application of automated external defibrillators (AEDs) along with high-quality CPR is crucial for favorable outcomes in patients with cardiac arrest. Here, we developed an automated CPR (A-CPR) apparatus that combines ACCD and AED and evaluated its performance in a pilot animal-based study.

Methods: Eleven pigs (n = 5, A-CPR group; n = 6, ACCD CPR and AED [conventional CPR (C-CPR)] group) were enrolled in this study. After 2 min observation without any treatment following ventricular fibrillation induction, CPR with a 30:2 compression/ventilation ratio was performed for 6 min, mimicking basic life support (BLS). A-CPR or C-CPR was applied immediately after BLS, and resuscitation including chest compression and defibrillation, was performed following a voice prompt from the A-CPR device or AED. Hemodynamic parameters, including aortic pressure, right atrial pressure, coronary perfusion pressure, carotid blood flow, and end-tidal carbon dioxide, were monitored during resuscitation. Time variables, including time to start rhythm analysis, time to charge, time to defibrillate, and time to subsequent chest compression, were also measured.

Results: There were no differences in baseline characteristics, except for arterial carbon dioxide pressure (39 in A-CPR vs. 33 in C-CPR, p = 0.034), between the two groups. There were no differences in hemodynamic parameters between the groups. However, time to charge (28.9 ± 5.6 s, A-CPR group; 47.2 ± 12.4 s, C-CPR group), time to defibrillate (29.1 ± 7.2 s, A-CPR group; 50.5 ± 12.3 s, C-CPR group), and time to subsequent chest compression (32.4 ± 6.3 s, A-CPR group; 56.3 ± 10.7 s, C-CPR group) were shorter in the A-CPR group than in the C-CPR group (p = 0.015, 0.034 and 0.02 respectively).

Conclusions: A-CPR can provide effective chest compressions and defibrillation, thereby shortening the time required for defibrillation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Animals, Laboratory
  • Carbon Dioxide
  • Cardiopulmonary Resuscitation*
  • Heart Arrest* / therapy
  • Pilot Projects
  • Swine

Substances

  • Carbon Dioxide

Grants and funding

This work was supported by the Korea Medical Device Development Fund grant funded by the Korea government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health & Welfare, the Ministry of Food and Drug Safety) (Project Number: 1465032025; 1465032980;1465036172). There was no additional external funding received for this study.