CT annulus sizing prior to transcatheter aortic valve replacement (TAVR): evaluation of free-breathing versus breath-holding acquisition

Eur Radiol. 2023 Dec;33(12):8521-8527. doi: 10.1007/s00330-023-09913-5. Epub 2023 Jul 20.

Abstract

Objectives: To assess feasibility and accuracy of aortic annulus measurements using cardiac computed tomography angiography (CTA) performed during free-breathing prior to transcatheter aortic valve replacement (TAVR).

Materials and methods: Sixty consecutive TAVR candidates underwent free-breathing wide-detector cardiac CTA, followed by a percutaneous valve replacement. For each, a theoretical valve size was suggested through CT measurements of the annulus, then compared to the size of the actual implanted transcatheter heart valve (THV). The procedural success and the 30-day outcomes were collected. Image quality of the annulus was also studied according to subjective and objective criteria. Data of a control group of 60 patients previously evaluated on breath-holding were also evaluated.

Results: A total of 120 patients (mean age, 83 years ± 7, 60 men) were evaluated. All CT acquisitions provided sufficient image quality allowing precise annulus measurements. Mean attenuation (p < 0.001) and image noise (p = 0.01) were higher in the free-breathing group, while image quality was comparable (p = 0.36). The agreement rate between CT-suggested valve size and THV implanted size was comparable, estimated at 87% (κ = 0.79, 95%CI 0.566, 0.908) on free-breathing vs. 82% (κ = 0.78, 95%CI 0.634, 0.904) on breath-holding. The procedure was successful for all patients without increase in 30-day mortality or adverse events.

Conclusions: Free-breathing cardiac CTA allows accurate aortic annulus measurements without compromising image quality or patients' outcome after TAVR. Elderly patients experiencing dyspnea, discomfort, or hearing loss that could prevent proper breath-holding should not be excluded from CT prior to TAVR.

Clinical relevance statement: To decrease elderly patients' discomfort, MDCT evaluation prior to transcatheter aortic valve replacement (TAVR) may be performed on quiet breathing with no significant impact on the outcome.

Key points: • Adhering to CT breathing commands can be challenging for patients with dyspnea, hearing impairment, agitation, or pulmonary diseases. • Free-breathing cardiac CT may be an alternative to breath-holding for patients unable to follow the breathing commands. • Wide-detector CT acquisition on free-breathing does not impair annulus measurements and prosthesis sizing in patients scheduled for TAVR.

Keywords: Aortic valve stenosis; Breath-holding; Multidetector computed tomography; Transcatheter aortic valve replacement.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / surgery
  • Computed Tomography Angiography
  • Dyspnea
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Multidetector Computed Tomography / methods
  • Predictive Value of Tests
  • Prosthesis Design
  • Transcatheter Aortic Valve Replacement* / methods
  • Treatment Outcome