Design and Application of Near Infrared LED and Solenoid Magnetic Field Instrument to Inactivate Pathogenic Bacteria

Micromachines (Basel). 2023 Apr 14;14(4):848. doi: 10.3390/mi14040848.

Abstract

Purpose: This study aims to evaluate the efficiency of infrared LEDs with a magnetic solenoid field in lowering the quantity of gram-positive Staphylococcus aureus and gram-negative Escherichia coli bacteria, as well as the best exposure period and energy dose for inactivating these bacteria.

Method: Research has been performed on a photodynamic therapy technique called photodynamic inactivation (PDI), which combines infrared LED light with a wavelength range of 951-952 nm and a solenoid magnetic field with a strength of 0-6 mT. The two, taken together, can potentially harm the target structure biologically. Infrared LED light and an AC-generated solenoid magnetic field are both applied to bacteria to measure the reduction in viability. Three different treatments infrared LED, solenoid magnetic field, and an amalgam of infrared LED and solenoid magnetic field, were used in this study. A factorial statistical ANOVA analysis was utilized in this investigation.

Results: The maximum bacterial production was produced by irradiating a surface for 60 min at a dosage of 0.593 J/cm2, according to the data. The combined use of infrared LEDs and a magnetic field solenoid resulted in the highest percentage of fatalities for Staphylococcus aureus, which was 94.43 s. The highest percentage of inactivation for Escherichia coli occurred in the combination treatment of infrared LEDs and a magnetic field solenoid, namely, 72.47 ± 5.06%. In contrast, S. aureus occurred in the combined treatment of infrared LEDs and a magnetic field solenoid, 94.43 ± 6.63 percent.

Conclusion: Staphylococcus aureus and Escherichia coli germs are inactivated using infrared illumination and the best solenoid magnetic fields. This is evidenced by the rise in the proportion of bacteria that died in treatment group III, which used a magnetic solenoid field and infrared LEDs to deliver a dosage of 0.593 J/cm2 over 60 min. According to the research findings, the magnetic field of the solenoid and the infrared LED field significantly impact the gram-positive bacteria S. aureus and the gram-negative bacteria E. coli.

Keywords: Escherichia coli; Infrared LED; Staphylococcus aureus; photoinactivation; solenoid magnetic field.