Spatial frequency-based correction of the spherical aberration in living brain imaging

Microscopy (Oxf). 2024 Feb 7;73(1):37-46. doi: 10.1093/jmicro/dfad035.

Abstract

Optical errors, including spherical aberrations, hinder high-resolution imaging of biological samples due to biochemical components and physical properties. We developed the Deep-C microscope system to achieve aberration-free images, employing a motorized correction collar and contrast-based calculations. However, current contrast-maximization techniques, such as the Brenner gradient method, inadequately assess specific frequency bands. The Peak-C method addresses this issue, but its arbitrary neighbor selection and susceptibility to the noise limit its effectiveness. In this paper, we emphasize the importance of a broad spatial frequency range for accurate spherical aberration correction and propose Peak-F. This spatial frequency-based system utilizes a fast Fourier transform as a bandpass filter. This approach overcomes Peak-C's limitations and comprehensively covers the low-frequency domain of image spatial frequencies.

Keywords: correction collar; objective lens; refractive index; spherical aberration.

MeSH terms

  • Brain*
  • Microscopy*
  • Neuroimaging