Comparative study between a gravity-based and peristaltic pump for intravenous infusion with respect to the generation of proteinaceous microparticles

Int J Pharm. 2023 Jul 25:642:123091. doi: 10.1016/j.ijpharm.2023.123091. Epub 2023 Jun 1.

Abstract

Subvisible particles generated during the preparation or administration of biopharmaceuticals might increase the risk of immunogenicity, inflammation, or organ dysfunction. To investigate the impact of an infusion system on the level of subvisible particles, we compared two types of infusion sets based on peristaltic movement (Medifusion DI-2000 pump) and a gravity-based infusion system (Accu-Drip) using intravenous immunoglobulin (IVIG) as a model drug. The peristaltic pump was found to be more susceptible to particle generation compared to the gravity infusion set owing to the stress generated due to constant peristaltic motion. Moreover, the 5-µm in-line filter integrated into the tubing of the gravity-based infusion set further contributed to the reduction of particles mostly in the range ≥ 10 µm. Furthermore, the filter was also able to maintain the particle level even after the pre-exposure of samples to silicone oil-lubricated syringes, drop shock, or agitation. Overall, this study suggests the need for the selection of an appropriate infusion set equipped with an in-line filter based on the sensitivity of the product.

Keywords: Flow imaging; Gravity infusion; Infusion pump; Peristaltic pump; Protein aggregation; Subvisible particle.