Purpose: To implement a pure power vector method for monocular subjective refraction using a regular phoropter with the only modification being the inclusion of a Stokes lens. The proposed methodology was tested with three different Stokes lenses, and the results were compared with conventional clinical refraction procedures.
Methods: Power vector subjective refraction was performed by attaching a Stokes lens to the Risley prism holder. Stokes lenses allow for pure astigmatic compensation in the form of the J0 , J45 components while the spherical lenses in the phoropter allow determination of the spherical component in the form of M (spherical equivalent). The proposed routine is presented step-by-step using three Stokes lenses having different astigmatic powers.
Results: Monocular subjective refraction was performed on 26 healthy subjects with a mean age of 44 ± 16 years, mean spherical equivalent of -0.56 D (range -5.50 to +2.38 D) and refractive astigmatism ≤1.50 D. No differences were found between the results obtained with the conventional technique versus the vector-based procedure for the spherical equivalent (p = 0.28) or astigmatic components (p = 0.34). In addition, visual acuity (VA) was equivalent through the refractions measured with the conventional and vector procedures (p = 0.12). Repeatability coefficients for J0 and J45 with the new vector methodology were <0.38 D.
Conclusions: The proposed routine could be helpful for cases where it is difficult to get a valid starting point for conventional refraction (e.g., irregular corneas and media opacities), for testing facilities with limited resources/equipment and/or for motivated clinicians who wish to know about alternative methods of refractive error determination.
Keywords: astigmatic decomposition; phoropter; power vectors; subjective refraction.