Radiation Exposure Reduction by Digital Variance Angiography in Lower Limb Angiography: A Randomized Controlled Trial

J Cardiovasc Dev Dis. 2023 Apr 30;10(5):198. doi: 10.3390/jcdd10050198.

Abstract

Background: digital variance angiography (DVA) provides higher image quality than digital subtraction angiography (DSA). This study investigates whether the quality reserve of DVA allows for radiation dose reduction during lower limb angiography (LLA), and compares the performance of two DVA algorithms.

Methods: this prospective block-randomized controlled study enrolled 114 peripheral arterial disease patients undergoing LLA into normal dose (ND, 1.2 µGy/frame, n = 57) or low-dose (LD, 0.36 µGy/frame, n = 57) groups. DSA images were generated in both groups, DVA1 and DVA2 images were generated in the LD group. Total and DSA-related radiation dose area product (DAP) were analyzed. Image quality was assessed on a 5-grade Likert scale by six readers.

Results: the total and DSA-related DAP were reduced by 38% and 61% in the LD group. The overall visual evaluation scores (median (IQR)) of LD-DSA (3.50 (1.17)) were significantly lower than the ND-DSA scores (3.83 (1.00), p < 0.001). There was no difference between ND-DSA and LD-DVA1 (3.83 (1.17)), but the LD-DVA2 scores were significantly higher (4.00 (0.83), p < 0.01). The difference between LD-DVA2 and LD-DVA1 was also significant (p < 0.001).

Conclusions: DVA significantly reduced the total and DSA-related radiation dose in LLA, without affecting the image quality. LD-DVA2 images outperformed LD-DVA1, therefore DVA2 might be especially beneficial in lower limb interventions.

Keywords: digital subtraction angiography; digital variance angiography; lower limb angiography; peripheral artery disease; radiation protection; radiation reduction.