Cranial Shape Measurements Obtained Using a Caliper and Elastic Bands Are Useful for Brachycephaly and Deformational Plagiocephaly Screening

J Clin Med. 2023 Apr 9;12(8):2787. doi: 10.3390/jcm12082787.

Abstract

We assessed a method for screening the cranial shape of 1-month-old infants using a simple measuring instrument instead of a three-dimensional scanner. The Mimos craniometer was used to measure cranial length, cranial width, and two diagonal lengths to calculate the cranial index (CI) and cranial asymmetry (CA). We defined a CI > 90% as brachycephaly and CA > 5 mm as deformational plagiocephaly (DP). Intra- and inter-examiner accuracy analyses were performed on a dummy doll and 1-month-old infants. The measurements of healthy 1-month-old infants were compared with previously reported three-dimensional scanner measurements. Intra- and inter-rater measurements showed good accuracy; diagnostic accuracy comparisons of brachycephaly and DP using a three-dimensional scanner showed kappa values of 1.0 and 0.8, respectively. Comparisons were made among 113 infants matched for day-age at the date of measurement; there were no significant differences in the CI (85.0% vs. 85.2%, p = 0.98) and CA (5.9 mm vs. 6.0 mm, p = 0.48) between the scanner and caliper measurements, nor in the prevalence of brachycephaly (12.4% vs. 17.7%, p = 0.35) or DP (58.4% vs. 56.6%, p = 0.89). This simple measurement method using calipers and bands was useful in screening for brachycephaly and DP in 1-month-old infants.

Keywords: brachycephaly; cranial shape; deformational plagiocephaly; infant; three-dimensional scanner.

Grants and funding

This research received no external funding.