Dual application of Polyvinyl Alcohol Glutaraldehyde Methylthymol Blue Fricke hydrogel in clinical practice: Surface dosimeter and bolus

Appl Radiat Isot. 2023 Jul:197:110827. doi: 10.1016/j.apradiso.2023.110827. Epub 2023 Apr 17.

Abstract

An essential issue is an accurate evaluation of surface dose distribution for such sensitive treatments. This work aimed to feasibility of the dual application of the Ferrous Polyvinyl Alcohol Glutaraldehyde Methylthymol Blue (PVA-GTA-MTB) gel as a bolus compensator and surface dosimeter in breast radiotherapy. The differences between the surface dose measured using PVA-GTA-MTB gel and film dosimetry in the medial and lateral parts of the breast were 3.74% and 4.18%, respectively. A qualitative comparison of the isodose curves showed that the PVA-GTA-MTB bolus creates a uniform dose distribution similar to the superflab bolus in the target volume.

Keywords: Bolus; Dosimetry; Film; Gel; Monte Carlo; Radiotherapy.

MeSH terms

  • Glutaral
  • Hydrogels
  • Phantoms, Imaging
  • Polyvinyl Alcohol*
  • Radiation Dosimeters
  • Radiometry* / methods
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods

Substances

  • polyvinyl alcohol hydrogel
  • Polyvinyl Alcohol
  • Glutaral
  • methylthymol blue
  • Hydrogels