Does an Extraoral Suction Device Reduce Aerosol Generation and Prevent Droplet Exposure to the Examiner during Esophagogastroduodenoscopy?

J Clin Med. 2023 Mar 29;12(7):2574. doi: 10.3390/jcm12072574.

Abstract

Esophagogastroduodenoscopy (EGD) is an aerosol-generating procedure. A major challenge in the COVID-19 era is how to prevent the spread of aerosols and droplets in endoscopic units. We evaluated the effectiveness of an extraoral suction device in preventing indoor aerosol diffusion and droplet exposure for examiners. The study involved 61 patients who underwent EGD at our institution from 1 February to 31 March 2022. To determine whether aerosol spread increases before or after EGD examination with an extraoral suction device located in front of the patient's mouth, aerosols of 0.3, 0.5, 1, 3, 5, and 10 μm were measured with a handheld particle counter. The degree of contamination of the plastic gowns on the examiners was assessed using the rapid adenosine triphosphate test. The extraoral suction device significantly reduced the diffusion of large particles (3, 5, and 10 μm) after finishing the EGD examination. However, the diffusion of small particles (0.3 and 0.5 μm) was significantly increased. This extraoral suction device was effective in reducing large particle diffusion during EGD examination but was limited for minimizing small particle diffusion or droplet exposure to the examiner.

Keywords: COVID-19; adenosine triphosphate hygiene monitoring test; aerosol; esophagogastroduodenoscopy; particle counter.

Grants and funding

This research is supported by a donated fund from laboratories established by Kagawa Prefecture.