Dosimetric investigation of whole-brain radiotherapy with helical intensity modulated radiation therapy and volumetric modulated arc therapy for scalp sparing

BJR Open. 2023 Mar 22;5(1):20220037. doi: 10.1259/bjro.20220037. eCollection 2023.

Abstract

Objective: Intensity-modulated radiotherapy (IMRT) is a well-established radiotherapy technique for delivering radiation to cancer with high conformity while sparing the surrounding normal tissue. Two main purposes of this study are: (1) to investigate dose calculation accuracy of helical IMRT (HIMRT) and volumetric-modulated arc therapy (VMAT) on surface region and (2) to evaluate the dosimetric efficacy of HIMRT and VMAT for scalp-sparing in whole brain radiotherapy (WBRT).

Methods: First, using a radiochromic film and water-equivalent phantom with three types of boluses (1, 3, 5 mm), calculation/measurement dose agreement at the surface region in the VMAT and HIMRT plans were examined. Then, HIMRT, 6MV-VMAT and 10MV-VMAT with scalp-sparing, and two conventional three-dimensional conformal radiotherapy plans (6MV-3DCRT and 10MV-3DCRT; as reference data) were created for 30 patients with brain metastasis (30 Gy/10 fractions). The mean dose to the scalp and the scalp volume receiving 24 and 30 Gy were compared.

Results: The percentage dose differences between the calculation and measurement were within 7%, except for the HIMRT plan at a depth of 1 mm. The averaged mean scalp doses [Gy], V24Gy [%], and V30Gy [%] (1SD) for 6MV-3DCRT, 10MV-3DCRT, HIMRT, 6MV-VMAT, and 10MV-VMAT were [26.6 (1.1), 86.4 (7.3), 13.2 (4.2)], [25.4 (1.0), 77.8 (7.5), 13.2 (4.2)], [23.2 (1.5), 42.8 (19.2), 0.2 (0.5)], [23.6 (1.6), 47.5 (17.9), 1.2 (1.8)], and [22.7 (1.7), 36.4 (17.6), 0.7 (1.1)], respectively.

Conclusion: Regarding the dose parameters, HIMRT achieved a lower scalp dose compared with 6MV-VMAT. However, the highest ability to reduce the mean scalp dose was showed in 10MV-VMAT.

Advances in knowledge: Scalp-sparing WBRT using HIMRT or VMAT may prevent radiation-induced alopecia in patients with BM.