Optimization of signal-to-noise ratio in short-duration SEP recordings by variation of stimulation rate

Clin Neurophysiol. 2023 Jun:150:89-97. doi: 10.1016/j.clinph.2023.03.008. Epub 2023 Mar 27.

Abstract

Objective: The intraoperative averaging of the somatosensory evoked potential (SEP) requires reliable recordings within the shortest possible duration. We here systematically optimized the repetition rate of stimulus presentation.

Methods: We recorded medianus and tibial nerve SEP during 22 surgeries and varied the rate of stimulus presentation between 2.7 Hz and 28.7 Hz. We randomly sampled a number of sweeps corresponding to recording durations up to 20 s and calculated the signal-to-noise ratio (SNR).

Results: For the medianus nerve at 5 s recording duration, SEP stimulation rate at 12.7 Hz obtained the highest median SNR = 22.9 for the N20, which was higher than for rate 4.7 Hz (p = 1.5e-4). When increasing the stimulation rate, latency increased and amplitude decayed for cortical but not for peripheral recording sites. For the tibial nerve, the rate 4.7 Hz achieved the highest SNR for all durations.

Conclusions: We determined the time-dependence of SNR for N20 and elucidated the underlying physiology. For short recordings, rapid reduction of noise through averaging at high stimulation rate outweighs the disadvantage of smaller amplitude.

Significance: For a short duration of medianus nerve SEP recording only, it may be advantageous to stimulate with a repetition rate of 12.7 Hz.

Keywords: Erb’s point; High frequency oscillation; Intraoperative neuromonitoring; Neurosurgery; Peripheral nerve conduction; Stimulation frequency.