The first mobile photon-counting detector CT: the human images and technical performance study

Phys Med Biol. 2023 Apr 25;68(9). doi: 10.1088/1361-6560/acc8b3.

Abstract

Objective. The purpose of this study is to assess its human images and its unique capabilities such as the 'on demand' higher spatial resolution and multi-spectral imaging of photon-counting-detector (PCD)-CT.Approach. In this study, the FDA 510(k) cleared mobile PCD-CT (OmniTom Elite) was used. To this end, we imaged internationally certified CT phantoms and a human cadaver head to evaluate the feasibility of high resolution (HR) and multi-energy imaging. We also demonstrate the performance of PCD-CT via first-in-human imaging by scanning three human volunteers.Main results. At the 5 mm slice thickness, routinely used in diagnostic head CT, the first human PCD-CT images were diagnostically equivalent to the EID-CT scanner. The HR acquisition mode of PCD-CT achieved a resolution of 11 line-pairs (lp)/cm as compared to 7 lp cm-1using the same kernel (posterior fossa-kernel) in the standard acquisition mode of EID-CT. For the quantitative multi-energy CT performance, the measured CT numbers in virtual mono-energetic images (VMI) of iodine inserts in the Gammex Multi-Energy CT phantom (model 1492, Sun Nuclear Corporation, USA) matched the manufacturer reference values with mean percent error of 3.25%. Multi-energy decomposition with PCD-CT demonstrated the separation and quantification of iodine, calcium, and water.Significance. PCD-CT can achieve multi-resolution acquisition modes without physically changing the CT detector. It can provide superior spatial resolution compared with the standard acquisition mode the conventional mobile EID-CT. Quantitative spectral capability of PCD-CT can provide accurate, simultaneous multi-energy images for material decomposition and VMI generation using a single exposure.

Keywords: PCD-CT; human images; photon-counting detector; techincal performance.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Head
  • Humans
  • Iodine*
  • Phantoms, Imaging
  • Photons*
  • Tomography Scanners, X-Ray Computed
  • Tomography, X-Ray Computed / methods

Substances

  • Iodine