Preliminary Study on an Alternative Test Method with MCTT HCETM for Ocular Irritation Test of Ophthalmic Medical Devices

Toxics. 2023 Mar 21;11(3):289. doi: 10.3390/toxics11030289.

Abstract

The sustained growth of the market for ophthalmic medical devices has increased the demand for alternatives to animal testing for the evaluation of eye irritation. The International Organization for Standardization has acknowledged the need to develop novel in vitro tests to replace animal testing. Here, we evaluated the applicability of an alternative method based on a human corneal model to test the safety of ophthalmic medical devices. 2-Hydroxyethyl methacrylate (HEMA) and Polymethyl methacrylate (PMMA), which are used to fabricate contact lenses, were used as base materials. These materials were blended with eye irritant and non-irritant chemicals specified in the OECD Test Guideline (TG) 492 and Globally Harmonized System (GHS) classification. Then, three GLP-certified laboratories performed three replicates using the developed method using 3D reconstructed human cornea epithelium, MCTT HCETM. OECD TG 492 describes the procedure used to evaluate the eye hazard potential of the test chemical based on its ability to induce cytotoxicity in a reconstructed human cornea-like epithelium (RhCE) tissue. Results: The within-laboratory reproducibility (WLR) and between-laboratory reproducibility (BLR) were both 100%. When a polar extraction solvent was used, the sensitivity, specificity, and accuracy were all 100% in each laboratory. When a non-polar extraction solvent was used, the sensitivity was 80%, the specificity was 100%, and the accuracy was 90%. The proposed method exhibited excellent reproducibility and predictive capacity within and between laboratories. Therefore, the proposed method using the MCTT HCETM model could be used to evaluate eye irritation caused by ophthalmic medical devices.

Keywords: alternative animal test; eye irritation test; medical devices; predictive capacity.