Quadruple UV LED Array for Facile, Portable, and Online Intrinsic Fluorescent Imaging of Protein in a Whole Gel Electrophoresis Chip

Anal Chem. 2023 Apr 18;95(15):6193-6197. doi: 10.1021/acs.analchem.2c05455. Epub 2023 Mar 28.

Abstract

Intrinsic fluorescence imaging (IFI) has been used for the stain-free detection of proteins in slab gel. However, complicated detection setups and small irradiation area limited the development of facile, online, and portable imaging of the whole slab gel. We here designed a quadruple UV LED array to produce even and powerful area light for direct irradiation of gel electrophoresis chip (GEC) at 275 nm. In addition, we only used a filter of 365 nm, a UV camera lens, and a CCD for IFI detection. We integrated the simple detection setup with the small GEC to construct the IFI-GEC device with a portable size of 15 × 15 × 38 cm. We detected three model proteins to demonstrate the good evenness of the LED array and the online imaging of the whole GEC. Furthermore, the reproducible IFI-GEC detection was completed within 10 min and the LOD was as low as 40 ng for lysozyme detection. All results indicated the potential of the IFI-GEC device for online and portable detection of proteins without staining.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrophoresis*
  • Optical Imaging / methods
  • Proteins* / analysis
  • Staining and Labeling

Substances

  • Proteins