Long-Term Outcomes of Articular Surface Replacement (ASR) Implant in Hip Arthroplasty: A Single Institution Review

Cureus. 2023 Mar 11;15(3):e36029. doi: 10.7759/cureus.36029. eCollection 2023 Mar.

Abstract

Various metal-on-metal (MoM) total hip replacements (THRs) have been found to have high short-term failure rates due to adverse responses to metal debris (ARMD). As a consequence, several low-performing THRs have been removed off the market. The purpose of this research was to look at the at least five-year outcomes of patients who had MoM hip arthroplasty at our institution. In one specialised centre between 2007 and 2008, 24 Articular Surface Replacement (ASRTM, DePuy, Warsaw, IN, USA) MoM THRs (in 24 patients, mean age: 56.4 years) were implanted. DePuy ASR hip prosthesis for osteoarthritis or hip fractures were employed in the THR system. All patients were summoned back for a clinical assessment, and imaging was done as needed. The average period of follow-up was 8.0 years (6.0-10 years). In all, eight instances (33.3%) were discovered to have pseudotumors, four hips (16.7%) were revised, and one (4.1%) was operated for ARMD. The Western Ontario and McMaster Universities Arthritis Index (WOMAC) and Oxford ratings improved statistically significantly five years after surgery in all three areas of pain, disability, and stiffness; however, there was no statistically significant change in the 36-Item Short Form Survey (SF-36) (mental) score. MoM hip arthroplasty had a greater revision incidence at five years in our group, presumably owing to the adoption of a smaller femoral head size.

Keywords: asian variant; depuy asr hip; long term follow up after orthopedic surgery; metal-on-metal total hips; total hip replacement (thr).