Dynamic Characterization of Thermocouples under Double-Pulse Laser-Induced Thermal Excitation

Sensors (Basel). 2023 Feb 21;23(5):2367. doi: 10.3390/s23052367.

Abstract

This study investigated the dynamic characteristics of thermocouples by using double-pulse laser excitation for dynamic temperature calibration under extreme conditions. An experimental device was constructed for double-pulse laser calibration; the device uses a digital pulse delay trigger to precisely control the double-pulse laser to achieve sub-microsecond dual temperature excitation with adjustable time intervals. The time constants of thermocouples under single-pulse laser excitation and double-pulse laser excitation were evaluated. In addition, the variation trends of thermocouple time constants under different double-pulse laser time intervals were analyzed. The experimental results indicated that the time constant increases and then decreases with the decrease in the time interval of the double-pulse laser. A method for dynamic temperature calibration was established for the evaluation of the dynamic characteristics of temperature sensors.

Keywords: double pulse laser; dynamic calibration; thermocouple; time constant.