Accurate Post-Calibration Predictions for Noninvasive Glucose Measurements in People Using Confocal Raman Spectroscopy

ACS Sens. 2023 Mar 24;8(3):1272-1279. doi: 10.1021/acssensors.2c02756. Epub 2023 Mar 6.

Abstract

In diabetes prevention and care, invasiveness of glucose measurement impedes efficient therapy and hampers the identification of people at risk. Lack of calibration stability in non-invasive technology has confined the field to short-term proof of principle. Addressing this challenge, we demonstrate the first practical use of a Raman-based and portable non-invasive glucose monitoring device used for at least 15 days following calibration. In a home-based clinical study involving 160 subjects with diabetes, the largest of its kind to our knowledge, we find that the measurement accuracy is insensitive to age, sex, and skin color. A subset of subjects with type 2 diabetes highlights promising real-life results with 99.8% of measurements within A + B zones in the consensus error grid and a mean absolute relative difference of 14.3%. By overcoming the problem of calibration stability, we remove the lingering uncertainty about the practical use of non-invasive glucose monitoring, boding a new, non-invasive era in diabetes monitoring.

Keywords: calibration stability; diabetes; in vivo Raman spectroscopy; multivariate data analysis; non-invasive glucose monitoring; portable sensor; tissue diagnostics.