Single Printing Step Prussian Blue Bulk-Modified Transducers for Oxidase-Based Biosensors

Biosensors (Basel). 2023 Feb 9;13(2):250. doi: 10.3390/bios13020250.

Abstract

We report on hydrogen peroxide sensors made through a single printing step with carbon ink containing catalytically synthesized Prussian blue nanoparticles. Despite their reduced sensitivity, the resulting bulk-modified sensors displayed both a wider linear calibration range (5 × 10-7-1 × 10-3 M) and an approximately four times lower detection limit versus the surface-modified sensors due to the dramatically decreased noise resulting in, on average, a six times higher signal-to-noise ratio. The corresponding glucose and lactate biosensors demonstrated similar and even higher sensitivities compared to those of biosensors based on surface-modified transducers. The biosensors have been validated through analysis of human serum. The decreased time and cost for production of single printing step bulk-modified transducers, as well as their analytical performance characteristics, which are advantageous over conventional surface-modified ones, would be expected to enable their wide use in (bio)sensorics.

Keywords: Prussian blue nanoparticles; biosensors; glucose; hydrogen peroxide; lactate; screen-printed electrodes.

MeSH terms

  • Biosensing Techniques* / methods
  • Electrodes
  • Humans
  • Oxidoreductases*
  • Printing, Three-Dimensional
  • Transducers

Substances

  • Oxidoreductases
  • ferric ferrocyanide