Use of threshold PEP with an adult non-rebreather oxygen mask plus prone positioning in acute hypoxemic respiratory failure due to SARS-CoV-2 infection during the collapse of the health system in a low-income country

SAGE Open Med Case Rep. 2023 Feb 15:11:2050313X231154064. doi: 10.1177/2050313X231154064. eCollection 2023.

Abstract

During the coronavirus disease 2019 pandemic, Ecuador reported a collapse of the healthcare system, in which intensive care unit beds were lacking. Therefore, we sought to determine whether the use of threshold expiratory positive pressure with an adult non-rebreather oxygen mask plus prone positioning is useful for improving oxygenation. Twelve patients were included. Eight patients (66.7%) survived, while four patients (33.3%) died. Baseline arterial oxygen saturation (%) prior placement median (interquartile range) 85.5% (80%-89%) and arterial oxygen saturation (%) post placement of the device was median (interquartile range) (93%-96%) (P = 0.0001). Respiratory rate before placement was median (interquartile range) 38 (36-42) and post placement of the device was median (interquartile range) 24 (22-30) (P = 0.0005). The use of an adapted device might be useful for the management of acute hypoxemic respiratory failure due to severe acute respiratory syndrome coronavirus 2 pneumonia, particularly when mechanical ventilators and high-flow oxygen systems are unavailable.

Keywords: COVID-19; adult; intensive care units; pandemics; prone position; respiratory distress syndrome; severe acute respiratory syndrome coronavirus 2.

Publication types

  • Case Reports