Design and ergonomic evaluation of a new adapted endoscope holder to reduce muscle pressure and discomfort during endoscopy

Work. 2023;76(1):403-410. doi: 10.3233/WOR-220185.

Abstract

Background: Laparoscopic surgeons who regularly perform endoscopy are more likely to develop musculoskeletal disorders than other internal medicine specialists, a difference that attributed to repetitive movements, poor postures, and sub-optimal equipment design.

Objective: This study aimed to design, build, and evaluate an endoscope holder for reducing the static load applied by the weight of the endoscope, in order to reduce musculoskeletal disorders risk factors in the surgeon's hand, shoulder and back issues regions.

Methods: A new endoscope holder was designed according to ergonomic design principles. The designed holder was evaluated by surface electromyography (sEMG) and discomfort assessment including 15 laparoscopic surgeons. The results were analyzed with centrality statistics and compared with the independent t-test using SPSS version 22.

Results: The evaluation of the new endoscope holder shows a statistical significant decrease in the average electrical activity of biceps brachii, triceps brachii, lateral deltoid, T9 Thoracic erector spinae, L4 Lumbar erector spinae, and external oblique after using the holder (p < 0.05).

Conclusion: The results shows that using the new endoscope holder is associated with a lower level of discomfort, as well as a lower induced muscle activity. The results also highlight the need to upgrade the holder to offer rotability in all directions (perpendicular to the ground) which will be included in the next design.

Keywords: EMG; Endoscope holder; co-design; discomfort; ergonomic design.

MeSH terms

  • Electromyography
  • Endoscopes
  • Endoscopy* / adverse effects
  • Ergonomics
  • Humans
  • Muscle, Skeletal / physiology
  • Musculoskeletal Diseases* / etiology
  • Musculoskeletal Diseases* / prevention & control