Air entrapment as a cause of early inappropriate shocks after subcutaneous defibrillator implant: A case series

Indian Pacing Electrophysiol J. 2023 May-Jun;23(3):84-87. doi: 10.1016/j.ipej.2023.02.001. Epub 2023 Feb 4.

Abstract

Inappropriate shock (IAS) is the most frequent device-related complication among patients with subcutaneous implantable cardioverter-defibrillator (S-ICD). Air entrapment (AE) has been described as an underdiagnosed cause of early postimplant IAS. We report 6 consecutive cases of early IAS after S-ICD implant, in whom the electrogram analysis (EGM) and/or chest radiography (CXR) were consistent with AE.

Keywords: Air entrapment; Complication; Inappropriate shock; S-ICD.