Electrochemical Aptasensor Developed Using Two-Electrode Setup and Three-Electrode Setup: Comprising Their Current Range in Context of Dengue Virus Determination

Biosensors (Basel). 2022 Dec 20;13(1):1. doi: 10.3390/bios13010001.

Abstract

We present, for the very first time, the fabrication and electrochemical characterization of a paper-based experimental platform for dengue virus analysis. The paper-based device incorporates a screen-printing technology with the help of black carbon conductive ink. The paper-based device utilizes two styles of electrode setups, i.e., the two-electrode system and three-electrode system, and both setups effectively detected the dengue virus with an LOD of 0.1 µg/mL; however, these paper electrodes exhibit various current ranges, and the created sensor was encompassed and compared in this research based on current response. It is observed that the three-electrode system has a substantially higher current range, ranging from 55.53 µA to 322.21 µA, as compared to the two-electrode system, which has a current range of 0.85 µA to 4.54 µA. According to this study, the three-electrode system displayed a good range of current amplification that is roughly 50 times higher than the two-electrode system, which had a weak current response. As a result, the three-electrode method has emerged as a viable option for the very sensitive detection of the dengue virus, as well as for the diagnosis of other diseases.

Keywords: comparative study; current amplification; dengue detection; paper electrode setups.

MeSH terms

  • Biosensing Techniques* / methods
  • Dengue Virus*
  • Electric Conductivity
  • Electrochemical Techniques / methods
  • Electrodes
  • Limit of Detection
  • Printing