Introduction: Viral load is a very useful marker for monitoring patients infected with HBV and HCV. This work compares assays based on transcription-mediated amplification (TMA) and on real-time PCR (RT-PCR) to verify whether they can be interchangeable.
Material and methods: A bicentric study, in which 147 plasma samples from patients infected with HBV and 229 with HCV were analyzed, was carried out. TMA-based assays (Aptima® HBV Quant and Aptima® HCV Quant Dx, employing Panther system (Hologic®)) and RT-PCR (COBAS® AmpliPrep/COBAS® TaqMan® and COBAS® 6800) were used and the degree of concordance between them was calculated.
Results: Viral load was detected in both systems in 60 (40.82%) HBV samples (median log viral load: COBAS: 2.51IU/mL (IQR 2.20-3.17), Panther: 2.71IU/mL (IQR 2.21-3.22)) and in 39 (16.96%) HCV samples (median log viral load: COBAS: 3.93IU/mL (IQR 2.24-6.01), Panther: 3.80IU/mL (IQR 1.99-6.14)). The agreement between both systems was κ=0.943 for HBV and κ=0.925 for HCV. Comparison of viral load samples detected by both assays showed a hight correlation for HBV (R2=0.86) and for HCV (R2=0.97).
Conclusions: Both TMA and RT-PCR based assays may be interchangeable for the management of patients infected with HBV and HCV.
Keywords: Aptima® Quant; HBV; HCV; On real-time PCR; PCR a tiempo real; Roche COBAS® 6800; Roche COBAS® AmpliPrep/COBAS®; TMA; TaqMan®; VHB; VHC.
Copyright © 2022 Sociedad Española de Enfermedades Infecciosas y Microbiología Clínica. Published by Elsevier España, S.L.U. All rights reserved.