Validation of SuPerSense, a Sensorized Surface for the Evaluation of Posture Perception in Supine Position

Sensors (Basel). 2022 Dec 30;23(1):424. doi: 10.3390/s23010424.

Abstract

This study aimed to validate a sensorized version of a perceptive surface that may be used for the early assessment of misperception of body midline representation in subjects with right stroke, even when they are not yet able to stand in an upright posture. This device, called SuPerSense, allows testing of the load distribution of the body weight on the back in a supine position. The device was tested in 15 patients with stroke, 15 age-matched healthy subjects, and 15 young healthy adults, assessing three parameters analogous to those conventionally extracted by a baropodometric platform in a standing posture. Subjects were hence tested on SuPerSense in a supine position and on a baropodometric platform in an upright posture in two different conditions: with open eyes and with closed eyes. Significant correlations were found between the lengths of the center of pressure path with the two devices in the open-eyes condition (R = 0.44, p = 0.002). The parameters extracted by SuPerSense were significantly different among groups only when patients were divided into those with right versus left brain damage. This last result is conceivably related to the role of the right hemisphere of the brain in the analysis of spatial information.

Keywords: body schema; center of pressure; load cells; spatial cognition; stabilometry.

MeSH terms

  • Adult
  • Body Weight
  • Humans
  • Perception
  • Posture*
  • Standing Position*
  • Supine Position