Validity of Spatio-Temporal Gait Parameters in Healthy Young Adults Using a Motion-Sensor-Based Gait Analysis System (ORPHE ANALYTICS) during Walking and Running

Sensors (Basel). 2022 Dec 28;23(1):331. doi: 10.3390/s23010331.

Abstract

Motion sensors are widely used for gait analysis. The validity of commercial gait analysis systems is of great interest because calculating position/angle-level gait parameters potentially produces an error in the integration process of the motion sensor data; moreover, the validity of ORPHE ANALYTICS, a motion-sensor-based gait analysis system, has not yet been examined. We examined the validity of the gait parameters calculated using ORPHE ANALYTICS relative to those calculated using conventional optical motion capture. Nine young adults performed gait tasks on a treadmill at speeds of 2−12 km/h. The three-dimensional position data and acceleration and angular velocity data of the feet were collected. The gait parameters were calculated from motion sensor data using ORPHE ANALYTICS, and optical motion capture data. Intraclass correlation coefficients [ICC(2,1)] were calculated for relative validities. Eight items, namely, stride duration, stride length, stride frequency, stride speed, vertical height, stance phase duration, swing phase duration, and sagittal angleIC exhibited excellent relative validities [ICC(2,1) > 0.9]. In contrast, sagittal angleTO and frontal angleIC demonstrated good [ICC(2,1) = 0.892−0.833] and moderate relative validity [ICC(2,1) = 0.566−0.627], respectively. ORPHE ANALYTICS was found to exhibit excellent relative validities for most gait parameters. These results suggest its feasibility for gait analysis outside the laboratory setting.

Keywords: foot-mounted sensor; foot-strike pattern; inertial measurement unit (IMU); reliability; validation study; wearable devices.

MeSH terms

  • Biomechanical Phenomena
  • Gait
  • Gait Analysis*
  • Humans
  • Reproducibility of Results
  • Running*
  • Spatio-Temporal Analysis
  • Walking
  • Young Adult