In Vivo Assessment and Monitoring of Burn Wounds Using a Handheld Terahertz Hyperspectral Scanner

Adv Photonics Res. 2022 May;3(5):2100095. doi: 10.1002/adpr.202100095. Epub 2022 Jan 17.

Abstract

The accuracy of clinical assessment techniques in diagnosing partial-thickness burn injuries has remained as low as 50-76%. Depending on the burn depth and environmental factors in the wound, such as reactive oxygen species, inflammation, and autophagy, partial-thickness burns can heal spontaneously or require surgical intervention. Herein, it is demonstrated that terahertz time-domain spectral imaging (THz-TDSI) is a promising tool for in vivo quantitative assessment and monitoring of partial-thickness burn injuries in large animals. We used a novel handheld THz-TDSI scanner to characterize burn injuries in a porcine scald model with histopathological controls. Statistical analysis (n= 40) indicates that the THz-TDSI modality can accurately differentiate between partial-thickness and full-thickness burn injuries (1-way ANOVA, p< 0.05). THz-TDSI has the potential to improve burn care outcomes by helping surgeons in making objective decisions for early excision of the wound.

Keywords: burn imaging; handheld THz-TDS scanner; histology; partial-thickness burn characterization; skin tissue hyperspectral imaging; terahertz time-domain spectroscopy.