Experimental Validation of Shifted Position-Diffuse Reflectance Imaging (SP-DRI) on Optical Phantoms

Sensors (Basel). 2022 Dec 15;22(24):9880. doi: 10.3390/s22249880.

Abstract

Numerous diseases such as hemorrhage, sepsis or cardiogenic shock induce a heterogeneous perfusion of the capillaries. To detect such alterations in the human blood flow pattern, diagnostic devices must provide an appropriately high spatial resolution. Shifted position-diffuse reflectance imaging (SP-DRI) has the potential to do so; it is an all-optical diagnostic technique. So far, SP-DRI has mainly been developed using Monte Carlo simulations. The present study is therefore validating this algorithm experimentally on realistic optical phantoms with thread structures down to 10 μm in diameter; a SP-DRI sensor prototype was developed and realized by means of additive manufacturing. SP-DRI turned out to be functional within this experimental framework. The position of the structures within the optical phantoms become clearly visible using SP-DRI, and the structure thickness is reflected as modulation in the SP-DRI signal amplitude; this performed well for a shift along the x axis as well as along the y axis. Moreover, SP-DRI successfully masked the pronounced influence of the illumination cone on the data. The algorithm showed significantly superior to a mere raw data inspection. Within the scope of the study, the constructive design of the SP-DRI sensor prototype is discussed and potential for improvement is explored.

Keywords: SP-DRI; all-optical; biomedical imaging; experimental validation; medical diagnostics; optical phantoms.

MeSH terms

  • Algorithms*
  • Diagnostic Imaging*
  • Humans
  • Models, Biological
  • Monte Carlo Method
  • Optical Imaging
  • Phantoms, Imaging