Enhancing Weak-Signal Extraction for Single-Molecule Localization Microscopy

J Phys Chem A. 2023 Jan 12;127(1):329-338. doi: 10.1021/acs.jpca.2c05164. Epub 2022 Dec 20.

Abstract

Single-molecule localization microscopy (SMLM) has been widely used in biological imaging due to its ultrahigh spatial resolution. However, due to the strategy of reducing photodamage to living cells, the fluorescence signals of emitters are usually weak and the detector noises become non-negligible, which leads to localization misalignments and signal losses, thus deteriorating the imaging capability of SMLM. Here, we propose an active modulation method to control the fluorescence of the probe emitters. It actually marks the emitters with artificial blinking character, which directly distinguishes weak signals from multiple detector noises. We demonstrated from simulations and experiments that this method improves the signal-to-noise ratio by about 10 dB over the non-modulated method and boosts the sensitivity of single-molecule localization down to -4 dB, which significantly reduces localization misalignments and signal losses in SMLM. This signal-noise decoupling strategy is generally applicable to the super-resolution system with versatile labeled probes to improve their imaging capability. We also showed its application to the densely labeled sample, showing its flexibility in super-resolution nanoscopy.

MeSH terms

  • Microscopy, Fluorescence / methods
  • Single Molecule Imaging* / methods