Acoustic Change Complex Recorded in Hybrid Cochlear Implant Users

Audiol Neurootol. 2023;28(3):151-157. doi: 10.1159/000527671. Epub 2022 Nov 30.

Abstract

Introduction: Expanding cochlear implant (CI) candidacy criteria and advances in electrode arrays and soft surgical techniques have increased the number of CI recipients who have residual low-frequency hearing. Objective measures such as obligatory cortical auditory-evoked potentials (CAEPs) may help clinicians make more tailored recommendations to recipients regarding optimal listening mode. As a step toward this goal, this study investigated how CAEPs measured from hybrid CI users differ in two listening modes: acoustic alone (A-alone) versus acoustic plus electric (A + E).

Methods: Eight successful hybrid CI users participated in this study. Two CAEPs, the P1-N1-P2 and the acoustic change complex (ACC), were measured simultaneously in response to the onset and change of a series of different and spectrally complex acoustic signals, in each of the two listening modes (A-alone and A + E). We examined the effects of listening mode and stimulus type on the onset and ACC N1-P2 amplitudes and peak latencies.

Results: ACC amplitudes in hybrid CI users significantly differed as a function of listening mode and stimulus type. ACC responses in A + E were larger than those in the A-alone mode. This was most evident for stimuli involving a change from low to high frequency.

Conclusions: Results of this study showed that the ACC varies as a function of listening mode and stimulus type. This finding suggests that the ACC can be used as a physiologic, objective measure of the benefit of hybrid CIs, potentially supporting clinicians in making clinical recommendations on individualized listening mode, or to document subjective preference for a given listening mode. Further research into this potential clinical application in a range of hybrid recipients and/or long electrode users who have residual low-frequency hearing is warranted.

Keywords: Acoustic change complex; Cochlear implant; Cortical auditory-evoked potential; Electric acoustic stimulation; Nucleus hybrid cochlear implant.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Acoustic Stimulation
  • Acoustics
  • Cochlear Implantation*
  • Cochlear Implants*
  • Evoked Potentials, Auditory / physiology
  • Hearing
  • Speech Perception* / physiology