Objective: The objectives of the study were to compare intraocular pressure (IOP) readings across a wide range and obtained via three rebound tonometers in ADAMTS10-mutant Beagle-derived dogs with different stages of open-angle glaucoma (OAG) and normal control dogs and to investigate the effect of central corneal thickness (CCT).
Animals studied: Measurements were performed on 99 eyes from 50 Beagle-derived dogs with variable genetics-16 non-glaucomatous and 34 with ADAMTS10-OAG. Seventeen OAG eyes were measured twice-with and without the use of IOP-lowering medications.
Procedures: IOP was measured in each eye using three tonometers with their "dog" setting-ICare® Tonovet (TV), ICare® Tonovet Plus® (TVP), and the novel Reichert® Tono-Vera® Vet (TVA)-in randomized order. CCT was measured with the Accutome® PachPen. Statistical analyses included one-way ANOVA, Tukey pairwise comparisons, and regression analyses of tonometer readings and pairwise IOP-CCT Pearson correlations (MiniTab®).
Results: A total of 116 IOP measurements were taken with each of the three tonometers. When comparing readings over a range of ~7-77 mmHg, mean IOPs from the TV were significantly lower compared with TVP (-4.6 mmHg, p < .001) and TVA (-3.7 mmHg, p = .001). We found no significant differences between TVA and TVP measurements (p = .695). There was a moderate positive correlation between CCT and IOP for TVA (r = 0.53, p < .001), TVP (r = 0.48, p < .001), and TV (r = 0.47, p < .001).
Conclusions: Our data demonstrate strong agreement between TVP and TVA, suggesting that the TVA may similarly reflect true IOP values in canines. CCT influenced IOP measurements of all three tonometers.
Keywords: ADAMTS10; canine; central corneal thickness (CCT); glaucoma; intraocular pressure (IOP); rebound tonometry.
© 2022 The Authors. Veterinary Ophthalmology published by Wiley Periodicals LLC on behalf of American College of Veterinary Ophthalmologists.