Mass photometry (MP) is a label-free, single-molecule technique that can determine molecular mass distribution with very low sample consumption in a short time. Because of the established experimental instrument and analytical software, MP measurements may be readily obtained; thus, the application of MP is expanding, especially in the fields of bioscience and biotechnology. However, because the MP data quality is strongly focus-dependent, optical settings must be intrinsically strict. In this study, we report the importance of the critical calibration of the mass photometer, which is required for the accurate estimation of high-molecular mass samples, such as adeno-associated virus vectors. Additionally, a method for optimizing the instrument settings, including the calibration of the stage, is presented.
Keywords: Adeno-associated virus vector; Gene therapy; Mass photometry.