Correlation between corneal and contact lens deformation with changes in intraocular pressure for wearable monitoring systems

Eye (Lond). 2023 Jul;37(10):2055-2060. doi: 10.1038/s41433-022-02285-y. Epub 2022 Oct 27.

Abstract

Objective: The aim of this work is to evaluate the extent to which the eye's curvature deformation, due to changes in the intraocular pressure (IOP), can be directly tracked by an overlying contact lens.

Method: In this experimental study, using 12 cadaveric eyes, the IOP was increased from 10 to 36 mmHg, while video imaging was used to capture the three experimental variations. The deformation of the bare eye was used as a control, while the deformation of an overlying silicone grided contact lens and an overlying microfluidic IOP-sensing contact lens were examined and compared.

Results: The relation between the slope of the radius of corneal curvature versus the IOP for both the bare eye and the marker contact lens yielded a linear relationship with a R2 value of 0.83. The microfluidic contact lens resulted in an average performance of 0.40 mm indicator movement/mmHg (SD 0.006). Comparing the slope of the marker contact lens deformation, to the performance of the microfluidic contact lens resulted in a R2 value of 0.78. The strain map of the overlaying grided contact lens showed most deformation occurring along the outer edge of the lens with increased deformation as increase IOP occurs; as well as with some negative, compressive movement near the central points.

Conclusion: The deformation from the curvature of the eye is significant enough from 10 to 36 mmHg that a silicone contact lens can capture and mimic those changes. The results show promise for optimization in contact lens-based IOP monitoring.

MeSH terms

  • Contact Lenses*
  • Corneal Diseases*
  • Intraocular Pressure
  • Silicones
  • Tonometry, Ocular
  • Wearable Electronic Devices*