Total aortic arch replacement using a frozen elephant trunk device: Results of a 1-year US multicenter trial

J Thorac Cardiovasc Surg. 2024 May;167(5):1680-1692.e2. doi: 10.1016/j.jtcvs.2022.08.029. Epub 2022 Sep 6.

Abstract

Objective: In this prospective US investigational device exemption trial, we assessed the safety and 1-year clinical outcomes of the Thoraflex Hybrid device (Terumo Aortic) for the frozen elephant trunk technique to repair the ascending aorta, aortic arch, and descending thoracic aorta.

Methods: For the trial, which involved 12 US sites, 65 patients without rupture were recruited into the primary study group, and 9 patients were recruited into the rupture group. All patients underwent open surgical repair of the ascending aorta, aortic arch, and descending thoracic aorta in cases of aneurysm and/or dissection. The primary end point was freedom from major adverse events (MAE), defined as permanent stroke, permanent paraplegia/paraparesis, unanticipated aortic-related reoperation (excluding reoperation for bleeding), or all-cause mortality.

Results: In the primary study group, 2 patients were lost to follow-up at 1 year. Freedom from MAE at 1 year was 81% (51/63). Seven patients (11%) died (including 2 before 30 days or discharge), 3 patients (5%) suffered permanent stroke, and 3 (5%) developed permanent paraplegia/paraparesis. Twenty-six patients (41%) underwent planned extension procedures, including 22 endovascular procedures within a median of 122 (interquartile range, 64-156) days. In the aortic rupture group, 2 patients were lost to follow-up at 1 year. Freedom from MAE at 1 year was 71% (5/7). One patient (14%) died, 2 patients (29%) had permanent stroke, and none had permanent paraplegia/paraparesis. No extension procedures were performed in the rupture group.

Conclusions: One-year results with the Thoraflex Hybrid device are acceptable. Long-term data are necessary to assess the durability of these repairs.

Keywords: aortic aneurysm; aortic arch (total); aortic dissection; frozen elephant trunk; hybrid repair.

Publication types

  • Clinical Trial
  • Multicenter Study

MeSH terms

  • Aorta, Thoracic / diagnostic imaging
  • Aorta, Thoracic / surgery
  • Aortic Aneurysm, Thoracic* / diagnostic imaging
  • Aortic Aneurysm, Thoracic* / etiology
  • Aortic Aneurysm, Thoracic* / surgery
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation*
  • Endovascular Procedures* / methods
  • Humans
  • Paraparesis / etiology
  • Paraplegia / etiology
  • Prospective Studies
  • Retrospective Studies
  • Stents
  • Stroke* / etiology
  • Treatment Outcome