Evaluations of the performances of PET and MRI in a simultaneous PET/MRI instrument for pre-clinical imaging

EJNMMI Phys. 2022 Oct 8;9(1):70. doi: 10.1186/s40658-022-00483-x.

Abstract

Background: PET/MRI is an attractive imaging modality due to the complementary nature of MRI and PET. Obtaining high quality small animal PET/MRI results is key for the translation of novel PET/MRI agents and techniques to the radiology clinic. To obtain high quality imaging results, a hybrid PET/MRI system requires additional considerations beyond the standard issues with separate PET and MRI systems. In particular, researchers must understand how their PET system affects the MR acquisitions and vice versa. Depending on the application, some of these effects may substantially influence image quality. Therefore, the goal of this report is to provide guidance, recommendations, and practical experiments for implementing and using a small animal PET/MRI instrument.

Results: Various PET and MR image quality parameters were tested with their respective modality alone and in the presence of both systems to determine how the combination of PET/MRI affects image quality. Corrections and calibrations were developed for many of these effects. While not all image characteristics were affected, some characteristics such as PET quantification, PET SNR, PET spatial resolution, PET partial volume effects, and MRI SNR were altered by the presence of both systems.

Conclusions: A full exploration of a new PET/MRI system before performing small animal PET/MRI studies is beneficial and necessary to ensure that the new instrument can produce highly accurate and precise PET/MR images.

Keywords: Fluorine-18; Gallium-68; Image quality; PET/MRI; Positron range; Quantification.